Sentinel V SC and RT Operation Manual
September
2017
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Page 189
Table 21:
Sentinel V Real-Time Input Command Summary
Command Default
Description
#ZS -01
Bin Size (cm) [60-480, -1=WS]
#ZV -01
Ambiguity Velocity (cm/s) [2-700,-1=WC/cos(angle)]
The # commands are for expert users, and changing them may have negative consequences
for a typical user!
Command Descriptions
Each listing includes the command’s purpose, format, default setting (if applicable) range, Recom-
mended Setting, and description. When appropriate, we include amplifying notes and examples. If a
numeric value follows the command, the Sentinel V Real-Time uses it to set a processing value (time,
range, percentage, processing flags). All measurement values are in metric units (mm, cm, and dm).
? – Help Menus
Purpose
Lists the major help groups.
Format
x? (see description)
Description Entering
?
by itself displays all command groups. To display help for one command group,
enter
x?
, where x is the command group to view. When the Sentinel V Real-Time displays
the help for a command group, it also shows the format and present setting of those com-
mands. To see the help or setting for one command, enter the command followed by a
question mark. For example, to view the CB command setting, enter
CB?
.
Examples
See below.
>==
CBREAK
SentinelV
Teledyne RD Instruments (c) 2016
All rights reserved.
Firmware Version: 66.01
>
>?
Available Commands:
A ----------------------- Sensor Commands
C ----------------------- Control Commands
E ----------------------- Environment Commands
F ----------------------- Fault Log Commands
O ----------------------- Feature Commands
P ----------------------- Performance Test Commands
T ----------------------- Time Commands
W ----------------------- Water Profiling Commands
Y ----------------------- Display Banner
Z ----------------------- Vertical Beam Profile Commands
? ----------------------- Display Main Menu>
>
>#?
*** CAUTION: These commands are reserved for RDI use
and may not be currently supported!
Available Commands:
#A ---------------------- Sensor Expert Commands