Calibration—492/492P Service Vol. 1 (SN B030000 & up)
Performance Check
Table 3-6
SPAN/DIV vs TIME MARKERS
FREQUENCY
SPAN/DIV
Time Mark Generator
Marker Output
20 MHz
50 ns
10 MHz
.1
μβ
5 MHz
.
2
/iS
2 MHz
.5 /xs
1 MHz
1
ms
500 kHz
2
MS
200 kHz
5
ms
100 kHz
1 0
ms
50 kHz
2 0
μβ
20 kHz
50
ms
10 kHz
.1
ms
5 kHz
.2 ms (Option 03)
2 kHz
.5 ms (Option 03)
1 kHz
1 ms (Option 03)
500 Hz
2 ms (Option 03)
11. Check Time/Div Accuracy (accuracy within 5%
of time selected)
a.
Test equipment setup is the same as that required for
step
1 0
.
b.
Apply the Marker Output from the time mark gener
ator directly to the RF INPUT and the Trigger output to the
492/492P EXT TRIG connector on the back panel. Set the
controls as follows:
FREQUENCY
«1 MHz
FREQ SPAN/DIV
100 kHz
RESOLUTION BANDWIDTH 10 kHz
REF LEVEL
-2 0 dBm
TIME/DIV
50 ms
Triggering
EXT
Min RF ATTEN
20 dB
c.
Apply 50 ms time markers. Tune the FREQUENCY
toward 0 Hz as the FREQ SPAN/DIV is reduced to zero, so
time markers are displayed on the time domain display (see
Fig. 3-9). Adjust FREQUENCY if necessary.
d.
Use the horizontal position control to align a marker
on the
1
st graticule line; then check the displacement of
markers from their respective positions over the center eight
divisions. Individual marker displacement must not exceed
5% or 2 minor divisions.
m. Remove the time mark generator signal and
reconnect the comb generator to the RF INPUT. Modulate
the 500 MHz comb generator with 2 *ts (500 kHz) markers
by applying the Marker Output to the Pulse Input of the
comb generator. Change the REF LEVEL to -2 0 dBm and
tune the FREQUENCY toward 500 MHz until 500 kHz
markers are displayed over the center eight divisions of
display.
n. Check the 500 kHz/div span accuracy.
o.
Reduce the RESOLUTION BANDWIDTH to 1 kHz,
then reduce the FREQ SPAN/DIV setting and time marks
(see Table 3-6) to check the accuracy for the 50 kHz
through 10 kHz FREQ SPAN/DIV selections.
p. For Option 03 instruments, reduce the RESOLUTION
BANDWIDTH to 100 Hz and set the TIME/DIV to 50 ms.
Check the 5 kHz through 500 Hz FREQ SPAN/DIV accura
cy. (This procedure will not yield the best amplitude display
but it will produce a readable frequency display.)
3odbh
r s n
- t - i
i
i.1
..
Position marker on graticule line
50Κ£/
.
,
^
« - ϊ
f c J « « a
.
F1" . 3-9. Display to illustrate how tim ing accuracy is checked.
3-22
REV AUG 1981
Содержание 492, 492P
Страница 12: ...492 492P Service Vol I SN B030000 up The 492 492P Spectrum Analyzer xii REV AUG 1981 ...
Страница 244: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up REV AUG 1981 5 81 ...
Страница 256: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 40 Frequency control encoder timing ...
Страница 263: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 43 9914 GPIA block diagram 5 100 REV AUG 1981 ...
Страница 299: ...Product 492 Ser 1 Date 2 12 85 Change Ref M55287 Fig 5 28 Basic tune voltage converter Page 2 of 6 ...