Maintenance—492/492P Service Vol. 1 (SN B030000 & up)
a. Test equipment required:
Spectrum analyzer
with tracking
generator.
Frequency Counter
Return Loss
Bridge
492/492P Spectrum
Analyzer with TR 503
tracking Generator; or
7L13 with a TR 502 Track
ing Generator.
TEKTRONIX DC 508
Wiltron Model 62BF50
c.
Connect the spectrum analyzer, tracking generator,
and frequency counter together as a system, with the fre
quency counter connected to the Auxiliary RF Output of the
tracking generator (see Fig. 4-14).
d.
Connect the spectrum analyzer/tracking generator
system through the return loss bridge to the Peltola jack
(J1029) on the 829 MHz amplifier board (see Fig. 4-14).
Reconnect P235 (100 MHz reference signal) and P237
(2182 MHz input) to the LO section of the converter.
b.
Unsolder and reconnect the jumper, on the 829 MHz
Amplifier board, to the test Peltola jack J1029 (see Fig.
4-13).
Terminate the 110 MHz Output (J232) connector with
50 Ω, using a bnc-to-Sealectro adapter and 20 dB bnc
attenuator. Pull the IF SELECT line high by switching to
band 2 (1.7—5.5 GHz).
/
Fig. 4-13. Location of te st jack and jum per on the 829 MHz Am plifier board.
REV JUN 1983
4-23
Содержание 492, 492P
Страница 12: ...492 492P Service Vol I SN B030000 up The 492 492P Spectrum Analyzer xii REV AUG 1981 ...
Страница 244: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up REV AUG 1981 5 81 ...
Страница 256: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 40 Frequency control encoder timing ...
Страница 263: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 43 9914 GPIA block diagram 5 100 REV AUG 1981 ...
Страница 299: ...Product 492 Ser 1 Date 2 12 85 Change Ref M55287 Fig 5 28 Basic tune voltage converter Page 2 of 6 ...