3.
Remove the nuts and lockwashers which hold the fan
brackets from the back side of the power supply housing.
The fan will fall free from the brackets.
4.
The resilient mounts from the corners of the fan frame
should be removed if a new fan is to be installed.
Replacement
1.
Insert 4 resilient mounts in the corners of the fan flush
with the fan frame.
2.
Install one of the fan brackets to the power supply
housing by attaching its lock washers and nuts to the back
of the housing.
NOTE
Fan brackets should be installed as in Fig. 4-9A.
3.
Insert the posts of the bracket into the holes provided
in the resilient mount, and install the remaining bracket with
the lockwashers and nuts to the back side of the power
supply housing.
•4. Reconnect the fan to the Fan Drive Board, then re
place the cover with the fan onto the power supply module.
5.
After installing the six screws that hold the cover in
place, ensure that the fan assembly moves freely. Replace
the coaxial cable in the plastic retaining clip.
6
.
Reinstall the power supply assembly as directed un
der Power Supply Replacement. Apply power and check for
normal fan operation.
Compliant Mounted Fan (SN B 0 3 0 0 0 0 to B042719)
Instruments SN B030000 to B042719 may have a com
pliant mounted fan (see Fig. 4-9B). It is important that the
mounting screws for these fans are not over-tightened.
Removal
1. Remove power supply as described in this section.
2.
Remove six screws that hold the power supply cover
in place. Take the coaxial cable out of the plastic retainer
clip and lift the power supply cover with fan up, so harmon
ica connector P3045 can be disconnected and the cover
removed.
3. Unsolder the leads to the fan.
Maintenance—492/492P Service Vol. 1 (SN B030000 & up)
4.
Using a 1/4 inch open-end wrench or needle-nose pli
ers, retain the nylon nut while unscrewing the fan mounting
screws with a Phillips screwdriver.
NOTE
After the fan is removed, be sure to retain the rubber
and steel washers on the screws. These are essential
for proper operation of the compliant mount. Do not
remove the self-locking nylon nuts.
5.
Remove the lead gasket from the old fan and install it
on the replacement fan assembly. Do not over-tighten the
screws retaining the gasket. The lead is soft enough to be
deformed by over-tightening.
Replacement
1. Refer to Fig. 4-9B and replace as follows:
a) place the foam gasket between the fan housing and
the power supply module cover;
b) insert the two 5/8 inch machine screws, with washers
mounted as shown in Fig. 4-9B, through the two eyelets.
Place the fan in position; then, while holding the nylon
nut with needle-nose pliers, screw the mounting screws
into the nut until the end of the screws is just flush with
the nut. Do not re-use these nuts because they can
loosen with vibration. The nylon nuts are self-locking.
2.
Re-solder the fan power leads, brown to pin 1, red to
pin 2, orange to pin 3, and yellow to pin 0.
3.
Reconnect the plug P3045, making sure the index
markers are aligned, then replace the cover with the fan
onto the power supply module.
4.
After installing the six screws that hold the cover in
place, ensure that the fan assembly moves freely. Replace
the coaxial cable in the plastic retaining clip.
5.
Reinstall the power supply assembly as directed un
der Power Supply Replacement. Apply power and check for
normal fan operation. If fan does not run, check the connec
tion of harmonica connector plug P3045.
REV JUN 1983
4-17
Содержание 492, 492P
Страница 12: ...492 492P Service Vol I SN B030000 up The 492 492P Spectrum Analyzer xii REV AUG 1981 ...
Страница 244: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up REV AUG 1981 5 81 ...
Страница 256: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 40 Frequency control encoder timing ...
Страница 263: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 43 9914 GPIA block diagram 5 100 REV AUG 1981 ...
Страница 299: ...Product 492 Ser 1 Date 2 12 85 Change Ref M55287 Fig 5 28 Basic tune voltage converter Page 2 of 6 ...