Calibration—492/492P Service Vol. 1 (SN B030000 & up)
Adjustment Procedure
k. Switch to the 10 dB/DIV mode and set the step
attenuators for 40 dB. Adjust Log Linearity, R1085 (Fig.
3-40), so the display is mid-screen.
I. If a large change in the setting of R1085 was required
in part k of this step, repeat the adjustments of R4071 and
R4081 because of interaction.
m. Check the accuracy of 10 dB/DIV and 2 dB/DIV dis
play modes by switching the attenuation in 10 dB steps for
10 dB/DIV mode and 1 dB steps for the 2 dB/DIV mode.
Note that the display steps one division ± 0.25 minor divi
sion for each 10 dB step, and ± 1.0 minor division for the
2 dB mode. Once the individual steps have been verified,
reset the signal level for full screen; then switch in the ap
propriate step attenuation to step the display down screen
to measure the worst case error over the dynamic range.
Maximum error must not exceed ±1.5 dB over the first
80 dB of range, or ±1.0 dB over the 16 dB range.
10. Calibrating the Resolution Bandwidth and
Shape Factor
NOTE
The filters are aligned separately and then combined
with a signal applied through both the VR#1 and
VR#2 modules. The final touch-up adjustments can
be made for filter shape and bandwidth. Because of
interaction, it is easy to offset one filter with
misadjustment of the other; therefore, only slight ad
justments should be made.
Adjust the bandwidth o f each filter section at the 3 dB
down level. This point should be as wide or slightly
wider than the 6 dB down point of the combined two
filter sections.
For gain levels and alignment theory, refer to
the
VR
part in the Theory of Operation section.
n. If the 10 dB log step in the 2 dB/DIV mode is long,
adjust gain with R4020 for less gain and rebalance R4071.
o.
Set the step attenuators for 10 dB of attenuation;
then adjust the signal output level for a full screen display
(+10 dBm) in the 2 dB/DIV mode.
a. Equipment setup is shown in Fig. 3-41.
1) place the VR module on an extender and connect the
output 10 MHz signal, from the 3rd Converter, to the in
put of the VR#2 section. Use a Sealectro male-to-male
adapter and coaxial cable to connect between P693 and
J683;
p. Activate MIN NOISE and switch out the 10 dB of
attenuation.
q. Check that the display level returns to within ± 1.0 dB
of reference.
r. Set the Ref Level to —15 dBm
and
adjust the signal
generator output for a full screen display in the 2 dB/DIV
mode.
s. Switch the Vertical Display to LIN and adjust Lin Bal,
R4052 (Fig. 3-40), for a full screen display. Display ampli
tude of LIN, 2 dB/DIV, and 10 dB/DIV display should now
be the same.
t. Check LIN mode linearity by adding
6
dB, 12 dB, and
18 dB of attenuation and noting that the display level is
down from top of screen four ( ± 0.4), six ( ± 0.4), and seven
(±0.4) divisions.
u.
Remove the signal generator signal connection to the
Log Amplifier input jack and replace P621.
2) connect the output of VR#2 to the input of the Log
Amplifier assembly by connecting a cable from J682 to
J621 (see Fig. 3-40);
3) connect the CAL OUT signal through a coaxial cable
to the RF INPUT;
4) set the front-panel controls as follows:
FREQUENCY
100 MHz
FREQ SPAN/DIV
50 kHz
RESOLUTION BANDWIDTH 100 kHz
REF LEVEL
- 2 0 dBm
b.
Tune the signal to center screen and change the Verti
cal Display to 2 dB/DIV. Adjust the REF LEVEL for a seven
division signal. Tune the display to center screen.
NOTE
The 10 kHz filter is used as the reference for centering
the response o f all filters.
REV AUG 1981
3-63
Содержание 492, 492P
Страница 12: ...492 492P Service Vol I SN B030000 up The 492 492P Spectrum Analyzer xii REV AUG 1981 ...
Страница 244: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up REV AUG 1981 5 81 ...
Страница 256: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 40 Frequency control encoder timing ...
Страница 263: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 43 9914 GPIA block diagram 5 100 REV AUG 1981 ...
Страница 299: ...Product 492 Ser 1 Date 2 12 85 Change Ref M55287 Fig 5 28 Basic tune voltage converter Page 2 of 6 ...