Rackmount/Benchtop Versions—492/492P Service Vol. 1 (SN B030000 & up)
Front rail t a p p e d fo r 1 0 * 3 2 scre ws
Left f r o n t
rail of
c a b i n e t ra c k
1 0 - 3 2
PHS
sc rew
(B) U n t a p p e d f r o n t rail
• N o t e :
Use FHS s c rew s if f r o n t rail h o l e s
a r e c o u n te r s u n k .
Fig. 6-4. Methods of mounting the stationary sections.
An alternative support method for the rear of the instru
ment is shown in Fig. 6-8. The rear support brackets, sup
plied with the instrument, are to be used in a rack which has
a spacing between the front and rear rails of 11 to 24
inches. Figure 6-8A illustrates the mounting method if the
rear rails are tapped for 10-32 screws and Fig. 6-8B illus
trates the mounting method if the rear rails are not tapped
for 10-32 screws.
R em oving o r Installing th e 492/4 9 2 P S p ectru m
A n aly z er from o r in th e R ack m o u n t C a b in e t
After the initial installation and adjustment of the slide-
out tracks, the 492/492P is removed or installed by follow
ing the directions in Fig. 6-6 and the following procedure:
1) remove the rackmount instrument from the rack and
place it on a bench;
If the rack does not have a rear rail, or if the distance
between the front and rear rails is too great, the instrument
may be mounted without the use of the slide-out tracks.
Fasten the instrument to the front rails of the rack with the
securing screws. This mounting method should be used only
if the instrument will not be subjected to shock or vibration
and it is installed in a stationary location.
S lide-out T rack L ubrication
The special finish on the sliding surfaces of the tracks
provide permanent lubrication. However, if the tracks do not
slide smoothly even after proper adjustment, a thin coating
of paraffin rubbed onto the sliding surfaces may improve
operation.
2) at the rear of the instrument, remove the Phillips screws
that hold the blower assembly on the back of the 492/492P
and pull the assembly off;
3) use a 5/32 Allen wrench to loosen and remove the four
screws that secure the rear panel of the 492/492P to the
back of the rackmount cabinet;
4) push or pull the instrument from the rackmount cabinet;
5) before the 492/492P will operate, a cut-off switch must be
closed. This is accomplished by reinstalling the blower as
sembly on the back panel of the 492/492P.
6-6
REV AUG 1981
Содержание 492, 492P
Страница 12: ...492 492P Service Vol I SN B030000 up The 492 492P Spectrum Analyzer xii REV AUG 1981 ...
Страница 244: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up REV AUG 1981 5 81 ...
Страница 256: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 40 Frequency control encoder timing ...
Страница 263: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 43 9914 GPIA block diagram 5 100 REV AUG 1981 ...
Страница 299: ...Product 492 Ser 1 Date 2 12 85 Change Ref M55287 Fig 5 28 Basic tune voltage converter Page 2 of 6 ...