I
I
I
I
I
h
I
i
I
I
I
I
I
I
I
I
I
I
d.
Repeat the procedure for 10kHz
bandwidth with a FREQ SPAN/DIV of 10 kHz.
resolution
e.
Change the PEAK/AVERAGE control to PEAK mode
and repeat the procedure for the 1 kHz resolution
bandwidth. (If the instrument has Option 03, reduce the
SPAN/DIV to 1 kHz.)
f.
If the 492/492P has Option 03, repeat the procedure
to check amplitude variation for resolution bandwidths of
100 Hz with a FREQ SPAN/DIV of 500 Hz. (Video Filter
must be off to maintain a calibrated display at 100 Hz
resolution.)
8. Check Frequency Response (± 1 .5 dB to
7.1 GHz, Bands 1, 2, and 3; and ±2.5 dB to 18 GHz,
Band 4)
Frequency response is the amplitude deviation, over a
given frequency range, of a constant level input signal mea
sured at the analyzer center frequency. It includes input at
tenuation, mixer and preselector (when installed) response,
plus mixing mode gain variations (band-to-band). Measure
ment requires many small incremental checks across the
spectrum analyzer frequency range. Response at each
check point must be optimized and separated from spurious
responses that are prevalent in instruments without the
preselector. Those instruments that have the preselector re
quire readjusting the PEAKING control at each check point
for frequencies above the range of band
1
(e.g., above
1.7 GHz). Because the frequency range of the 492/492P is
very wide, measuring the response in small increments is a
slow process. A more expeditious method using a sweep
oscillator is described in this procedure.
The procedure for checking frequency response depends
on the 492/492P configuration (options installed). Proce
dures for each configuration are described in four parts.
Re
fer to the appropriate part o f this step to check your
instrument.
Test equipment (see Table 3-4) is the same for
each procedure. If your instrument is the rackmount version
with semi-rigid cables to the back panel (Option 31), fre
quency response may degrade at the higher frequency end
(see Rackmount/Benchtop Version, Section
6
for details).
NOTE
Loss o f signal through interconnecting cables be
comes significant above 1 GHz; therefore, use short
(25 inch or less) semi-rigid cable with precision fittings
to interconnect the test equiment. Precision matching
terminations and power dividers are used to minimize
reflections.
Table 3-4
RECOMMENDED TEST EQUIPMENT
Calibration—492/492P Service Vol. 1 (SN B030000 & up)
Performance Check
Equipment
Recommended
Signal Generator
10 kHz— 10 MHz
HP654
Signal Generator
10 MHz— 18.0 GHz
HP8620C Sweeper with
86222B and HP86290A
RF Plug-in Units
Power Meter
0 to —10 dBm,
10 MHz— 18GHz
HP435A with 8481A
and 8482A Sensors
Power Divider
HP1167A
3 dB Attenuator,
SMA Connectors
Weinschel Model 4M
High Performance 50 Ω
Cable, SMA Connectors
See Equipment Required
list
50 Ω Coaxial Cable,
BNC Connectors
See Equipment Required
list
Adapter
N male-to-SMA male
See Equipment Required
list
Part 1
Procedure for Instruments Without Options 01 or 02
(Preselector and Digital Storage)
a.
Test equipment setup is shown in Fig. 3-4. Set the
front-panel controls as follows:
FREQUENCY RANGE
FREQ SPAN/DIV
FREQUENCY
MIN RF ATTEN
REF LEVEL
AUTO RESOLUTION
TIME/DIV
Vertical Display
0—4.2 GHz
(0— 1.8 GHz Option 01)
1 MHz
5 MHz
30 dB
0 dBm
On
2 0
ms
10 dB/DIV
REV AUG 1981
b.
Apply the output of a 100 kHz to 10 MHz signal gen
erator, with an output monitor, to the RF INPUT of the
492/492P. Set the generator frequency to 5 MHz and its
output for a
— 6
dBm to —10 dBm signal level.
c.
Change the Vertical Display to 2 dB/Div and adjust
the 492/492P REF LEVEL so the amplitude of the signal
generator signal is about half screen.
3-15
Содержание 492, 492P
Страница 12: ...492 492P Service Vol I SN B030000 up The 492 492P Spectrum Analyzer xii REV AUG 1981 ...
Страница 244: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up REV AUG 1981 5 81 ...
Страница 256: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 40 Frequency control encoder timing ...
Страница 263: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 43 9914 GPIA block diagram 5 100 REV AUG 1981 ...
Страница 299: ...Product 492 Ser 1 Date 2 12 85 Change Ref M55287 Fig 5 28 Basic tune voltage converter Page 2 of 6 ...