Maintenance—492/492P Service Vol. 1 (SN B030000 & up)
Removal
1. Remove the front-panel assembly as described
previously.
2. Use an Allen wrench to loosen the knob locking screw
and remove all knobs.
3.
Lay the front-panel on its face; then remove the elev
en circuit board screws and the screw that heat sinks and
holds IC U6090 on the board. Note that the screw next to
the connector plug has a fiber washer.
4. To prevent losing the grounding rings or bushings be
tween the front-panel controls and the front-panel casting,
hold the circuit board against the front-panel casting while
turning the complete assembly so it rests on the base of the
crt assembly.
5. Gently lift the casting from the circuit board. Ensure
that the grounding rings remain on the shaft of all controls
as the casting is removed.
Replacement
Reverse the removal procedure, ensuring that the fiber
washer is on the board screw next to the connector plug.
This washer prevents the screw from shorting a circuit
board run to the front-panel casting.
Replacing Front-Panel Push Button Switches
The front-panel assembly does not have to be removed
to replace any push button switch. The procedure follows:
1.
Remove front-panel knobs. Loosen and remove nuts
and washers for the RF INPUT, EXTERNAL MIXER, plus
the 1 st and 2nd LO connectors.
2.
Remove the screw that was under the FREQUENCY
tuning knob which holds the panel to the front-panel casting.
3.
Loosen the black screws through the crt bezel so the
panel can be moved enough to lift it off the casting.
4. Unplug and replace the desired switches.
Main Power Supply Module
To avoid damage to the Mother board connector
J5041 and Interface connector J 1034 during removal
or installation of the Power Supply Module, use the
following procedure.
Removal
1.
Disconnect the power cord, set the 492/492P on its
face or front panel and remove the instrument cover.
2.
Unplug the coaxial cable connector P620 from the
Log Amplifier assembly and pull the cable through so it is
clear.
3. Remove the three screws that hold the power module
to the RF deck flange (bottom right side).
4. Remove the four screws that hold the power supply
module to the side-rails.
5.
With the instrument on its face and the RF deck on
the near side, pull the left side of the power module from its
side-rail (no more than one and one-half inch). Now grasp
both sides of the module and lift to separate the module
from the Mother board.
W A R N I N G
I
Because C6011 and C6101 discharge very slowly,
hazardous potentials exist within the power supply for
several minutes after the power switch is turned off. A
relaxation oscillator, formed by C5113, R5111, and
DS5112, indicate the presence of voltages in the cir
cuit until the potential across the filter capacitors is
below 80 V.
Replacement
1.
Set the instrument face down with the RF deck on the
near side.
REV AUG 1981
4-15
Содержание 492, 492P
Страница 12: ...492 492P Service Vol I SN B030000 up The 492 492P Spectrum Analyzer xii REV AUG 1981 ...
Страница 244: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up REV AUG 1981 5 81 ...
Страница 256: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 40 Frequency control encoder timing ...
Страница 263: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 43 9914 GPIA block diagram 5 100 REV AUG 1981 ...
Страница 299: ...Product 492 Ser 1 Date 2 12 85 Change Ref M55287 Fig 5 28 Basic tune voltage converter Page 2 of 6 ...