i
i
i
i
f t
*
)
d.
Check the remaining TIME/DIV selections for ± 5% or
less error over the center eight divisions.
e.
Set the TIME/DIV to AUTO, FREQ SPAN/DIV to
MAX, and activate AUTO RESOLUTION.
Table 3-11
RESOLUTION AND SWEEP RATE
AS A FUNCTION OF SPAN IN THE AUTO MODE
Calibration—492/492P Service Vol. 1 (SN B030000 & up)
Adjustment Procedure
f.
Check the Time/Div versus Resolution Bandwidth as
per Table 3-11 for the different FREQ SPAN/DIV settings.
g.
Return Triggering to FREE RUN and remove the time-
mark generator markers to the 492/492P. Reposition the
trace if moved in part c.
Table 3-11 (cont)
FREQ SPAN/DIV
RESOLUTION
TIME/DIV
FREQ SPAN/DIV
RESOLUTION
TIME/DIV
MAX
1 MHz
50 ms
200 kHz
10 kHz
50 ms
200 MHz
1 MHz
2 0
ms
100 kHz
10 kHz
2 0
ms
100 MHz
1 MHz
1 0
ms
50 kHz
10 kHz
1 0
ms
50 MHz
1 MHz
1 0
ms
20 kHz
10 kHz
1 0
ms
20 MHz
1 MHz
1 0
ms
10 kHz
1 kHz
50 ms
10 MHz
1 MHz
1 0
ms
OPTION 03 ONLY
5 MHz
1 MHz
1 0
ms
5 kHz
1 kHz
2 0
ms
2 MHz
100 kHz
1 0
ms
2 kHz
1 kHz
1 0
ms
1 MHz
100 kHz
1 0
ms
1 kHz
100 Hz
,5 s
500 kHz
100 kHz
1 0
ms
500 Hz
100 Hz
1
s
■
■
Fig. 3-29. Location of timing adjustment R5105 and TP 1061 on Sweep board.
REV JUN 1983
3-49
Содержание 492, 492P
Страница 12: ...492 492P Service Vol I SN B030000 up The 492 492P Spectrum Analyzer xii REV AUG 1981 ...
Страница 244: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up REV AUG 1981 5 81 ...
Страница 256: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 40 Frequency control encoder timing ...
Страница 263: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 43 9914 GPIA block diagram 5 100 REV AUG 1981 ...
Страница 299: ...Product 492 Ser 1 Date 2 12 85 Change Ref M55287 Fig 5 28 Basic tune voltage converter Page 2 of 6 ...