Maintenance—492/492P Service Vol. 1 (SN B030000 & up)
Exceptions for Versions 1.1 and 1.2 in the 492P
only
The INC and DEC arguments for the TIME command do
not operate if the display is uncalibrated (UNCAL indicator
lit). Also, if 0 or a negative number is used as an argument
for the TIME command, no error will be reported and the
command is not executed.
Changes incorporated in Version 1.2 firmware:
1. INIT resets the display data pointer to its power-up
value: 500,225.
2.
Pressing RESET TO LOCAL while a message, includ
ing a REPEAT command, is executing, limits message ex
ecution to 256 times, if the message contains WAIT. A
SIGSWP command preceding WAIT in the message is ig
nored after RESET TO LOCAL button is pressed. The RE
PEAT loop, therefore, completes faster.
3.
The command DELFR OFF is inserted after FINE OFF
at the beginning of SET query response. The DELFR re
sponse is also sent after the FREQ response.
These changes remove the uncertainty in how the SET
query is executed with various combinations of instrument
settings.
TROUBLESHOOTING ON THE INSTRUMENT BUS
Instrument Bus Data Transfers
The 492P can execute two commands and queries to aid troubleshooting of circuit functions that are interfaced to the
instrument bus. These functions are configured by data sent from the microcomputer or respond with data for the
microcomputer: in either case, the data is transferred over the instrument bus. The commands and queries provided for this
purpose are:
Command
Query
Action
ADDR
ADDR?
Sets, returns address for DATA command
DATA
DATA?
Sets, returns data on 492P instrument bus
Because the DATA command can change the status of internal functions, its use may prevent normal operation of the
492P.
Using a 4050-Series controller, these commands and queries are transmitted to the 492P with the PRINT statement:
LINE «
P R I
n t
GPIB PRIMARY
ADDRESS
192P MESSAGE
3784-159
4-42
REV JUN 1983
Содержание 492, 492P
Страница 12: ...492 492P Service Vol I SN B030000 up The 492 492P Spectrum Analyzer xii REV AUG 1981 ...
Страница 244: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up REV AUG 1981 5 81 ...
Страница 256: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 40 Frequency control encoder timing ...
Страница 263: ...Theory of Operation 492 492P Service Vol 1 SN B030000 up Fig 5 43 9914 GPIA block diagram 5 100 REV AUG 1981 ...
Страница 299: ...Product 492 Ser 1 Date 2 12 85 Change Ref M55287 Fig 5 28 Basic tune voltage converter Page 2 of 6 ...