EUD-2016026-00 Rev.2, SmartScanIS User’s Guide for Union Pacific, 14 January 2021
35
The figure below shows the control panel that's on the front of a calibrated heat source
(2100-810NG).
3.2.1 Temperature Knob
The temperature knob allows you to select one of five delta temperatures. The temperature
label on each knob setting represents degrees Fahrenheit above ambient. The
180
setting is
normally used with all STC scanners.
3.2.2 Gating Switch
On the front of the heat source is the
Gating
switch. Toggling it on causes gating signals to
be sent to the Function I/O connectors. The first signal sent is a simulated TO1 (aka
transducer-A) signal. The second signal sent is a simulated TO2 (aka transducer-B) signal.
As long as this switch is toggled on, this sequence continues at a rate of 10 signals per second
(five of them being TO1s and five of them being TO2s).
The gating signals aren't needed for calibration of the SmartScanIS. The
Gating
switch may
be left in the off position.
FUNCTION I/O
130° BOLO
SOUTHERN TECHNOLOGIES CORP.
220°
126°
180°
DEG. (F)
130° PYRO
LO
2100-810NG
GATING
OFF
ON
HI
FUSE
TEMPERATURE
2 AMP
Содержание SmartScanIS
Страница 1: ...EUD 2016026 00 Rev 2 SmartScanIS User s Guide for Union Pacific 14 January 2021...
Страница 82: ...82 14 January 2021 SmartScanIS User s Guide for Union Pacific EUD 2016026 00 Rev 2...
Страница 90: ...90 14 January 2021 SmartScanIS User s Guide for Union Pacific EUD 2016026 00 Rev 2...
Страница 130: ...130 14 January 2021 SmartScanIS User s Guide for Union Pacific EUD 2016026 00 Rev 2...
Страница 258: ...258 14 January 2021 SmartScanIS User s Guide for Union Pacific EUD 2016026 00 Rev 2...
Страница 376: ...376 14 January 2021 SmartScanIS User s Guide for Union Pacific EUD 2016026 00 Rev 2...
Страница 380: ...380 14 January 2021 SmartScanIS User s Guide for Union Pacific EUD 2016026 00 Rev 2...