13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
13.2.1.5
Adjustment of the ceph primary diaphragm
13.2.1.5.1
Automatic adjustment: Ceph - Primary diaphragm (standard)
Phantom for Ceph exposure
✔
The needle phantom is
not
in the bite block holder on the unit.
✔
The ceph test phantom connects to the cover of the sensor
✔
The ceph sensor is plugged into the ceph slot on the unit.
✔
The ear plug holders on the ceph arm have been moved completely
apart and swung out of the beam direction (ap).
1. In the structure tree, under
"Ceph"
, click on the
"Adjustment of Ceph
primary diaphragm"
element (S010.3).
The
"Adjustment of Ceph primary diaphragm"
menu is displayed
in the action area.
2. Establish [ → 209] receptivity.
NOTICE
Risk of damage to unit
It is essential that the needle phantom is removed from the bite block
holder on the unit before a ceph exposure is taken; otherwise, the
sensor could collide with the phantom.