64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
163
Sirona Dental Systems GmbH
11 Commissioning, demo mode and repacking
Installation Manual ORTHOPHOS SL
11.12 Service routines for commissioning and repacking
båÖäáëÜ
11.12.6
S032: Sensor test
11.12.6.1
S032: Test step 10
Self-test of the DCS image data path
1. Call service routine S032.10 [ → 133].
Once the service routine has been selected, "DCS test" is
displayed in selection field 1.
2. Press the T key.
The DCS image data path is tested.
If the image data path is OK, "Success" is displayed in the
selection field 1.
3. Exit the service routine [ → 136].
SR*
SA**
Function
S032
Sensor test
S032.10
No
Self-test of the DCS image data path
S032.50
No
Self-test of the FP image data path