13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the calibration menu
Installation Manual ORTHOPHOS SL
3. Create an exposure (80 kV / 14 mA; 14.9 s) [ → 209].
Adjustment value S1 is calculated automatically from the
exposure and entered in the text box of the
"Adjustment of Ceph
main beam direction"
menu.