11 Commissioning, demo mode and repacking
Sirona Dental Systems GmbH
11.12 Service routines for commissioning and repacking
Installation Manual ORTHOPHOS SL
11.12.4.7.1
Codes of possible error modes
You can draw conclusions concerning the correctness of the angle
acquisition based on the plausibility of these displays. A plausible
constellation (reference unit) would be:
+22° → AD = 1273
-17° → AD = 3124
The settings may differ unit specifically.
Display
Error
Troubleshooting measures
01
Normal state
No action
02
Plug X1000 is not plugged in on
board DX1
Check plug for occlusal bite block on board DX1.
03
Persistent ADC values
Check cabling and plug connections on board DX1 and occlusal
bite block electronics.
04
Interruption of the 5 V line
05
Interruption of V
ref-
line
06
Interruption of V
out-
line
07
Interruption of V
GND
line
08
Interruption of "Connect" line
09
Defective choke, pin 3 detached
Replace occlusal bite block electronics (board DX51).
10
Defective choke, pin 1 detached
11
Defective clock generator
12
Resistor R11 not soldered
13
Plug on occlusal bite block not
inserted
Check plug for occlusal bite block on occlusal bite block
electronics (board DX51).
14
Composite error
Check cabling and plug connections on board DX1 and occlusal
bite block electronics.
15
Calculated angle too small
Perform a readjustment via service routine S018.9.
16
Calculated angle too large
17
Error on ADC channel 2
Replace occlusal bite block electronics (board DX51); check
cabling and plug connections on board DX1 and occlusal bite
block electronics.