13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
13.2.3
Resetting adjustment/calibration
In an absolute emergency, the settings made for adjustment and
calibration can be reset in full or in part to the factory settings and/or
modified manually in the
"Adjustment/calibration"
menu.
1. To reset adjustment and calibration data, click in the structure tree
under
"Reset"
on the element
"Reset Pan"
,
"Reset Ceph"
or
"Reset
3D"
(depending on what data you would like to reset).
2. To start resetting the adjustment and calibration data, click on the
"Reset"
button.
After resetting the values
If the adjustment and calibration has been reset, the unit must be
readjusted and recalibrated.
NOTICE
Create a data backup of the adjustment/calibration data in all cases
before you reset the adjustment and calibration.
IMPORTANT
With the
"Reset 3D"
function, only the
"Sensor adjustment"
and
"Diaphragm adjustment"
adjustments are reset. The calibration values
from the
"Sensor calibration"
,
"Diaphragm calibration"
and
"Geometry
calibration"
menus are retained.