64 95 142 D3632
250
D3632
.
031.01.02
.
02 03.2016
13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the calibration menu
Installation Manual ORTHOPHOS SL
13.2.1.7.2
Manual adjustment: Ceph secondary diaphragm (QuickShot)
The manual adjustment procedure is similar to the one for automatic
adjustment [ → 247]. The only difference is that the default adjustment
values calculated automatically in the
"Adjustment of Ceph secondary
diaphragm (Quickshot)"
menu are overwritten by adjustment values
calculated manually.
1. Start the
"Adjustment of Ceph secondary diaphragm (Quickshot)"
adjustment procedure as described in the chapter entitled "Automatic