64 95 142 D3632
266
D3632
.
031.01.02
.
02 03.2016
13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the calibration menu
Installation Manual ORTHOPHOS SL
13.2.2.4
Diaphragm configuration
✔
The 3D calibration block
is not
in the bite block holder on the unit.
1. In the structure tree, under
"3D"
/
"Diaphragm calibration"
, click on
the
"5x5"
element (S030.11).
The
"5x5"
volume is selected.
2. Establish [ → 209] receptivity.
3. Create an exposure (85 kV / 0.34 mAs) [ → 209].
A message is displayed in the action window to indicate whether
adjustment has been successful and the status indicator in front
of the
"5x5"
element in the structure tree responds accordingly:
Calibration not OK = status indicator stays red
Calibration OK = status indicator turns green
NOTE: The calibration values are automatically saved.
4. Click on the
"Next"
button or select the volume
"8x8"
in the structure
tree (S030.12).
The
"8x8"
volume is selected.
5. Establish [ → 209] receptivity.