64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
247
Sirona Dental Systems GmbH
13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
13.2.1.7
Adjustment of the ceph secondary diaphragm (QuickShot)
13.2.1.7.1
Automatic adjustment: Ceph secondary diaphragm (QuickShot)
(standard)
Phantom for Ceph exposure
✔
The pan needle phantom is
not
in the pan bite block holder on the
unit.
✔
The ceph test phantom does
not
connect to the cover of the sensor
unit.
1. In the structure tree, under
"Ceph"
, click on the
"Adjustment of Ceph
secondary diaphragm (Quickshot)"
element (S010.8).
The
"Adjustment of Ceph secondary diaphragm (Quickshot)"
menu is displayed in the action area.
2. Establish [ → 209] receptivity.
NOTICE
Risk of damage to unit
It is essential that the needle phantom is removed from the bite block
holder on the unit before a ceph exposure is taken; otherwise, the
sensor could collide with the phantom.