11 Commissioning, demo mode and repacking
Installation Manual ORTHOPHOS SL
11.12 Service routines for commissioning and repacking
båÖäáëÜ
11.12.4.3
S018: Test step 4
Check the height adjustment sensor system
This service routine is used to move the unit up or down as far as the limit
switches using the Up/Down keys on the control panel. The "soft limit
positions" set by the software are ignored in this case.
1. Call service routine S018.4.
Once the service routine has been selected, the current height
position is displayed in selection field 1.
Patient symbol keys 1 to 3 (F) show the switching state of the limit
switches (see table)
If the RESPECTIVE patient symbol key is lit, the corresponding
switch is activated.
2. Use the UP/DOWN keys on the control panel to move the unit up and
down and use the patient symbol keys (F) to check the switching
states.
3. Exit the service routine.
Display on the control
panel
Status
Meaning
Patient symbol key 1
lit
Correction switch activated
not lit
Correction switch not
activated
Patient symbol key 2
lit
Lower limit switch activated
not lit
Lower limit switch not
activated
Patient symbol key 3
lit
Upper limit switch activated
not lit
Upper limit switch not
activated