13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.3 Adjusting the touchscreen via the web service
Installation Manual ORTHOPHOS SL
13.3
Adjusting the touchscreen via the web service
✔
The unit and computer are turned on.
✔
The unit is logged into the network as an X-ray component.
1. Start an Internet browser such as Internet Explorer or Firefox.
2. In the address line, enter "https://" and the IP address of your unit.
Example: https://192.168.15.240
Tip: The IP address of your unit can be found on the info screen.
3. Confirm your input with the "Enter" key.
The Sirona Web service is loaded.
The sub menus of the
"User"
menu are displayed in the structure
tree.
4. Click on the
"Easypad Touchcalibration"
button.
The service routine for touchscreen calibration is opened.
You can abort the adjustment procedure at any time by clicking
on the
"Cancel"
button in the Web browser.
5. Click on the
"Display Easypad Touchcalibration Screen"
button.