13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.1 General information on unit adjustment and calibration
båÖäáëÜ
13.1.3
Enabling exposure readiness
To take an exposure, the system must first be made ready for exposure.
✔
The calibration menu must be opened [ → 195].
✔
Select the corresponding element in the structure tree.
➢ Click the
"Start acquisition"
button in the calibration menu.
Exposure readiness will be established. A dialog window displays
the status of readiness to exposure.
The service routine used for the corresponding exposure is
displayed on the control panel, along with the specific exposure
parameters.
13.1.4
Taking an exposure
✔
The calibration menu must be opened.
✔
Select the corresponding element in the structure tree.
✔
The unit must be made ready for exposure [ → 209].
➢ Press the release button. Press and hold down the button until the
exposure is complete, the X-ray image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if configured [ → 146]).
13.1.5
Save values
✔
The adjustment and calibration must have been performed correctly.
➢ To save the adjustment and calibration values, click the
"Validate and
save"
button in the calibration menu.
The adjustment and calibration is saved.
The saved adjustment and calibration is identified in the structure
tree by a check mark (adjustment) or a green traffic light
(calibration).