11 Commissioning, demo mode and repacking
Installation Manual ORTHOPHOS SL
11.12 Service routines for commissioning and repacking
båÖäáëÜ
11.12.6.3
Explanations on the test procedure
The synchronized readout sequence of the sensor is performed with real
data transmissions and CAN bus commands.
The image data are analyzed in the memory of the DX11.
It is possible to evaluate the LEDs on DX1 (image, pan) during the
synchronized readout sequence.
Image LED V501
Pan LED V500
In addition, "wiggle tests" of cables or ring rotation tests can also be
performed.
LED is lit
Synchronized readout sequence
active
LED not lit
Standby mode
LED is lit
Sensor unit available / standby mode
LED not lit
Sensor unit not available / standby mode
LED not lit
Sensor unit available / synchronized readout
sequence active