![Sirona ORTHOPHOS SL 2D Скачать руководство пользователя страница 97](http://html1.mh-extra.com/html/sirona/orthophos-sl-2d/orthophos-sl-2d_installation-manual_1285015097.webp)
64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
97
Sirona Dental Systems GmbH
10 Installation: Ceph arm
Installation Manual ORTHOPHOS SL
10.3 Installation
båÖäáëÜ
Installing the ceph arm
1. Attach the support arm to the four bearing bolts.
NOTICE! The connecting cables must not be crushed. Make sure
that the connecting cables lie in the groove of the support arm and
are not crushed.
2. Screw in the conical nuts (J) and tighten them firmly.
3. Detach the cables from the holder. Lay the cables over the support
arm in the groove between the slide and the support arm (from left to
right toward the opening (K)).
4. NOTICE! Never crush the cables. Please ensure that the cover can
be attached easily. Reattach the cover (B).
5. Remove the styrofoam packaging from the ceph arm.
6. Follow the instructions in the section entitled “Installing the secondary
IMPORTANT
The support arm may get tilted while you are tightening the conical nuts.
You can avoid this by lifting and moving the support arm gently while you
are tightening the nuts.
4x
J
B