13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
13.2.1.8
Adjustment of the ceph main X-ray beam
13.2.1.8.1
Automatic adjustment: Ceph main X-ray beam direction (standard)
✔
The ceph test phantom connects to the unit's sensor unit [ → 212].
✔
The ear plug holders on the ceph arm have been swung out of the
beam direction (ap).
1. In the structure tree, under
"Ceph"
, click on the
"Adjustment of Ceph
main beam direction"
element (S010.6).
The
"Adjustment of Ceph main beam direction"
menu is
displayed in the action area.
2. Establish [ → 209] receptivity.
NOTICE
Risk of damage to unit
It is essential that the needle phantom is removed from the pan bite
block holder on the unit before a ceph exposure is taken; otherwise, the
sensor unit could collide with the phantom.