13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.1 General information on unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.1.6.2
Adjustment phantom for adjustment of the cephalometer
For
"Adjustment of Ceph primary diaphragm"
and
"Adjustment of Ceph
main beam direction"
ceph adjustments, the ceph adjustment phantom
(A) must be inserted into the sensor unit cover.
For the
"Adjustment of Ceph secondary diaphragm"
ceph adjustment,
the ceph adjustment phantom must be removed from the sensor unit
cover once again.
A