13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
2. Align the calibration block with the spirit level by adjusting the screw
(A).
* For greater clarity, the unit in the left image is displayed without
sensor unit.
3. Open the calibration menu [ → 195].
4. In the structure tree, under
"3D"
, click on the
"Sensor adjustment"
element (S010.20).
5. Establish readiness for exposure [ → 209].
6. Create an exposure (85 kV / 22 mA s) [ → 209].
A message is displayed in the action window to indicate whether
adjustment has been successful and the values can be applied.
7. Repeat the procedure starting at Step 5 until adjustment is
successful.
8. When adjustment is OK, save the values [ → 209].
A message is displayed in the action window to indicate that the
values have been successfully saved.
The status indicator that prefixes the element
"Sensor
adjustment"
is checked.
9. Remove the 3D calibration block from the pan bite block holder on the
unit.
10. Continue with the next stage of the adjustment procedure [ → 264].
*