8.2 Connecting the line voltage
båÖäáëÜ
1. Check the protective ground wires and the unit leakage current
according to IEC 62353: 2007 (see sections “Checking the protective
ground wires [ → 84]” and “Checking the leakage current of the
2. Record the measured values in the document titled “Inspection and
maintanance and safety-related checks” (REF 64 95 100).
3. Now connect the power supply as shown.
8.2.2
Connecting the media converter
➢ Insert the media converter's power supply connector into the plug