64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
227
Sirona Dental Systems GmbH
13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
13.2.1.3
Pan symmetry adjustment
13.2.1.3.1
Automatic adjustment: Pan - Symmetry
✔
The needle phantom is in the bite block holder on the unit [ → 210].
1. In the structure tree, under
"Pan"
, click on the
"Symmetrie
adjustment"
element (S010.2).
The
"Pan - Symmetrie"
menu is displayed in the action area.
2. Establish [ → 209] receptivity.
3. Create an exposure (60 kV / 3 mA; 14.1 s) [ → 209].
The adjustment values S1, S2 and S3 are automatically
determined from the exposure and entered in the text boxes of
the
"Symmetrie adjustment"
menu.
The exposure is displayed in the exposure window.