64 95 142 D3632
172
D3632
.
031.01.02
.
02 03.2016
11 Commissioning, demo mode and repacking
Sirona Dental Systems GmbH
11.12 Service routines for commissioning and repacking
Installation Manual ORTHOPHOS SL
1. Call service routine S037.4.
Once the service routine has been selected, the IP address of the
unit is displayed in selection field 1.
"DEFAULT", or "STATIC" or "DYNAMIC" is displayed in selection
field 2 (see table).
2. You can display various items of network data in selection field 1 by
pressing the patient symbol keys (A, B, or C) (see table).
The patient symbol key selected in each case lights up.
3. To change the selected parameter, first press the T key (E).
4. Now use the patient symbol keys to select the required number pad
1-4 (A-D) (see also table):
A = Number pad B1
B = Number pad B2
C = Number pad B3
D = Number pad B4
The patient symbol key selected in each case lights up.
The digit currently selected for changing is displayed in selection
field 1 ("Digit No. 3" in the example).
Important: The number of the digit always refers to the currently
selected number pad.
The current value of the corresponding digit is displayed in
selection field 2 ("2" in the example).