11 Commissioning, demo mode and repacking
Sirona Dental Systems GmbH
11.12 Service routines for commissioning and repacking
Installation Manual ORTHOPHOS SL
11.12.3.4
S017: Test step 6
Activate / deactivate the remote control display
* factory settings
1. Call service routine S017.6.
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
2. Use the arrow keys (A) to select the code for the required setting in
selection field 1 (see table).
Once the required setting has been selected, the Memory key (B)
lights up.
3. Save the setting.
4. Exit the service routine.
Selection field
Code
Function
1
00
Display inactive*
01
Display active