64 95 142 D3632
268
D3632
.
031.01.02
.
02 03.2016
13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the calibration menu
Installation Manual ORTHOPHOS SL
13.2.2.5
Geometry calibration
1. Insert the 3D calibration block into the bite block holder on the
2. Align the 3D calibration block with the spirit level by adjusting the
screw (A).
* For greater clarity, the unit in the left image is displayed without
sensor unit.
3. In the structure tree, under
"3D"
/
"Geometry calibration"
/
"5x5"
,
click on the
"Molar left"
/ element (S010.25 - S010.39).
The
"Molar left"
element is selected.
4. Establish [ → 209] receptivity.
5. Create an exposure (85 kV / 22 mA s) [ → 209].
A message is displayed in the action window to indicate whether
adjustment has been successful and the status indicator in front
of the
"5x5"
/
"Molar left"
element in the structure tree responds
accordingly:
Calibration not OK = status indicator stays red
Calibration OK = status indicator turns green
NOTE: The calibration values are automatically saved.
6. Repeat the procedure starting at Step 4 until calibration is successful.
7. Click on the
"Next"
button or select the next center of rotation of
volume
"5x5"
.
*
A