13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the calibration menu
Installation Manual ORTHOPHOS SL
6. If adjustment is OK (the adjustment values in the text boxes are equal
to zero), save the values. [ → 209]
The status indicator that prefixes the
"Sensor adjustment"
element is checked.
7. Remove the needle phantom from the bite block holder of the unit.
8. Continue with the next stage of the adjustment procedure. [ → 221]