Sirona Dental Systems GmbH
11 Commissioning, demo mode and repacking
Installation Manual ORTHOPHOS SL
11.11 Using demo mode – operation without radiation release
båÖäáëÜ
11.11
Using demo mode – operation without radiation
release
If the unit is to be presented as a demo unit at trade fairs or exhibitions, it
must be ensured that radiation release is blocked.
11.11.1
Switching on demo mode
When operated in demo mode, the unit must not release any radiation.
For this reason, you must take the following safety measures:
1. Turn the device off.
2. Remove the "Rear tube assembly" cover.
3. Remove the cover plate of board DX6.
4. Set dip switch S2 (DX6) to position 2.
IMPORTANT: If switch S2 is not set to position 2 in demo mode
before switching off the unit, various error messages will display
when the unit is turned back on.
5. Pull cable L5 (XRAY) off connector J6 / J103 (DX6).
Radiation release is now no longer possible.
6. Switch on the unit and check the mode on the info screen (see
Operating Instructions).
Demo mode: ON means that: Demo mode is switched on (radiation
release is not possible)
Demo mode: OFF means that: Demo mode is switched off
(radiography, X-ray radiation are possible!)
7. Switch the unit off again and reattach the cover plate and the tube
assembly cover by following the dismantling procedure in reverse
order.
DANGER
Perilous shock hazard!
It is essential to switch off the unit and to wait at least another 4 minutes
before taking off the covers of the X-ray tube assembly.