Examination
Fluoroscopy
AXIOM Iconos R200
AXD3-340.620.13.01.02
11
/ 52
Fluoroscopy time limit
Fluoroscopy is switched off automatically and can no longer be switched on after
a fluoroscopy time of 10 minutes
1
.
To continue working
◆
let the fluoroscopy switch go and press it once again.
Fluoroscopy warning signal
An acoustic signal sounds after a fluoroscopy time of maximum 5 minutes
2
.
◆
The signal disappears when you press the key on the Fluorospot Compact.
The time for the acoustic signal can be reset to zero at any time.
Automatic format collimation in fluoroscopy
Automatic format collimation is also switched on automatically when fluoroscopy
is released.
I.e. the fluoroscopic field is never larger than the selected image intensifier for-
mat.
You can limit the fluoroscopic field to the required size with the corresponding
operating elements within this area.
Fluoroscopy programs
All parameters necessary for fluoroscopy are stored in programs.
The following programs can be stored:
– Standard fluoroscopy
– Supervision
– Pulsed fluoroscopy (Carevision)
– Roadmap standard fluoroscopy
– Roadmap Supervision
– Roadmap pulsed fluoroscopy
➩
Refer to the Fluorospot Compact Operator Manual.
1
Can be changed on request by SIEMENS Uptime Service, under compliance with the national
regulations.
2
Can be configured from 1 to 5 minutes by the SIEMENS Uptime Service
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...