Technical Description
Technical Data
AXIOM Iconos R200
AXD3-340.620.13.01.02
7
/ 12
Cassette spotfilm de-
vice (optional)
Cassettes:
Front loading for cassettes standardized according to IEC, ANSI and DIN (metal
or plastic) of the formats 18 cm x 24 cm (8” x 10”) up to 35 cm x 43 cm (14” x
17”). For permissible sizes and segmentations refer to cassette program;
automatic loading, centering and format sensing
Automatic format
collimation:
Can be selected separately according to format height/width for cassette spot-
filming; automatic formatting for Bucky exposures, object-related collimation pos-
sible after 1 to 2 seconds
Film segmentation:
Max. 4 on 1; secondary width diaphragms
Inward movement time:
Park to exposure position max. 1s (24 cm x 30 cm cassette, without lead inlay)
Transition time:
Fluoroscopy - radiographt 1s ± 15% (without filter positioning/grid movement);
1.35 s ± 15% with oscillating grid
Series technique:
With multiple segmentation and DFR
Tabletop-film distance:
7.8 cm (minimum)
Tabletop-I.I. distance:
11.8 cm (minimum)
I.I. image receptor/cas-
sette spotfilm device
(optional)
Scattered radiation grid:
Stationary 17:1, 70 lines/cm, f
0
= 125 cm;
Oscillation for CR cassettes configurable by Uptime Service;
Motorized grid movement in or out of beam path in 4 s
Spotfilm device travel:
Max. 105 cm (remote-controlled), fine-adjustable up to max. 8 cm/s (routine) or
16 cm/s (Periscanning);
PERISTEPPING option: stepping with max. 5 steps of 18 cm each (I.I. 40)
Central ray height above
the floor:
77 cm to 182 cm with table tilted up (+ 90°);
Tabletop-image intensi-
fier screen distance:
7.8 cm (minimum, I.I. image receptor only)
Distance from central
beam to end of table:
minimum 42 cm (head end)
Attenuation of the
useful radiation
Attenuation factor without cone m = 2,48
Attenuation factor with cone m = 2,50
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...