Examination
Digital Radiography
AXIOM Iconos R200
AXD3-340.620.13.01.02
43
/ 52
If the LED in the start key flashes,
◆
press the key
– The compression device moves into park position, if present.
– Oblique setting moves into the 0° position.
Once these positions are reached, the flashing LED goes out.
◆
Select an organ program with serialography according the type of examina-
tion.
Moving the system and tabletop into the starting position
◆
Position the patient.
◆
Bring the X-ray system and tabletop into the intended starting position.
The exposure series is generally started in the pelvic region for better exposure
control.
The entire examination area must be scanned without interruption. If necessary,
perform a test run without fluoroscopy and correct the starting position.
◆
Deflect the joystick for “X-ray system longitudinal movement” in the direction
of the head until the entire examination area has been scanned.
◆
Return to the starting position.
◆
Check the starting position with fluoroscopy switched on.
Scanning process
The X-ray system is in the starting position for the scanning process.
◆
Press the exposure release button until the entire examination area has been
scanned.
– The X-ray system starts moving automatically when the exposure is
released.
End of the scene
◆
Let go of the exposure release button.
– The X-ray system movement stops automatically
Start
During the exposure series the patient should take only shallow breaths or not
breathe at all.
The scene is also ended if the image memory is full and the maximum scene
time has been reached, independently of the position in which the system is lo-
cated.
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...