
System Operation
On-Off/Emergency Stop
6
/ 32
AXD3-340.620.13.01.02
Operator Manual
Switching the system off
◆
Press the button on the generator ON/OFF console.
– The entire system and all further connected devices are immediately
switched off.
– All started operating sequences will be interrupted and the selected
programs will be deselected.
➩
Refer to the FLUOROSPOT Compact Operator Manual for further information.
Emergency SHUTDOWN button (installed
on-site)
Use this shutdown method
only in an extreme emergency
because it is an
un-
controlled
process!
Data could be lost, e.g. unsaved images, exporting and filming jobs, etc. The tube
cooling system is also disconnected from the power supply, and the tube can
overheat.
Standby power supply
Only continuous fluoroscopy is possible in standby operation.
Pulsed fluoroscopy and all radiography modes are not possible.
“Standby power supply active” appears on the integrated generator control con-
sole in the message line.
Standby power supply inside the hospital
A standby power supply (inside the hospital) switches on if the line voltage fails.
However, this is not as a rule without interruption and can take some time.
The system is switched off and must be switched on anew.
Battery-supported standby power supply (optional)
UPS = Uninterrupted Power Supply
An uninterrupted power supply is a battery-supported system for the standby
power supply in the case of mains power supply failures.
If an UPS is installed for your system, then this UPS switches automatically to
standby operation if the mains voltage fails or is too low. To prevent loss of data/
images, the imaging system remains switched on. The generator and table are
switched off.
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...