AXIOM Iconos R200
AXD3-340.620.13.01.02
7
/ 52
Examination
Fluoroscopy
Fluoroscopy operating modes
❏
Standard fluoroscopy (continuous fluoroscopy)
– In continuous fluoroscopy 30 (25)
1
frames per second are generated by the
television central control unit.
❏
Fluoroscopy with pulsed television scanning (Supervision option)
– In SUPERVISION only 15 (12.5)
1
frames per second are generated. In this
case fluoroscopy is applied at half the radiation dose with half the temporal
resolution.
or
❏
Primary pulsed fluoroscopy (CAREVISION option)
– In pulsed fluoroscopy, the radiation is pulsed at 15 (12.5)
1
or 7.5 (8.3)
1
or 3
(3)
1
pulses per second. In this way a corresponding dose saving with a
lower temporal resolution is possible.
❏
Roadmap (only with DSA option) with standard fluoroscopy, with Supervision
(optional) or pulsed fluoroscopy (optional)
– Roadmap is a special fluoroscopic mode with subtraction display
Selecting the fluoroscopy operating mode
All parameters necessary for fluoroscopy are stored in organ programs.
◆
Select a corresponding organ program on the integrated generator control
console.
Changing the selection of the fluoroscopy
operating mode
It is possible to change the selection of the operating modes in connection with
the Supervision or CAREVISION options.
The current operating mode is displayed in the operating menu of the live moni-
tor.
Use the FLUOROSPOT Compact infrared control for operation.
If your system is equipped only with standard fluoroscopy, the and Fluo-
ro- keys have no function.
1
Values in brackets apply to VIDEOMED DHC
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...