System Operation
Setting the Image Geometry
30
/ 32
AXD3-340.620.13.01.02
Operator Manual
Additional Cu filter
The organ programs determine which additional Cu filter is moved into the beam
path on selection of the program.
A temporary change of the automatic additional Cu filter selection with organ pro-
gram is possible by manual switchover. The manually selected additional Cu filter
is maintained until renewed switchover or selection of an organ program with
configured additional Cu filter.
The currently selected additional Cu filter is displayed on the control units and pri-
mary collimator.
Image reversal
You can mirror the images on the image monitor horizontally or vertically with
these 2 keys. If none of the keys is pressed, then the image is displayed in the
normal position.
Image mirrored horizontally / vertically
Warning
It is part of the user’s duty to pay attention to the correct
orientation of the display on the monitor. Use suitable aids for this,
e.g. Pb letters.
Electronically displayed right/left markings are not mirrored when the image is
reversed.
A faulty diagnosis is possible due to incorrect position information, e.g. confu-
sion between right and left, with legal consequences and under certain circum-
stances with severe consequences for the patient.
R
R
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...