Examination
Digital Radiography
AXIOM Iconos R200
AXD3-340.620.13.01.02
35
/ 52
Programming run
Proceeding from the abdomen (starting position), the primary collimators, semi-
transparent wedge filters and dominant, if required, are set step by step (up to
the last position).
You must deflect the joystick or press the key for “X-ray system longitudinal
movement” in the movement direction for each step as long as the next step po-
sition is reached. Once the next step position is reached, the X-ray system stops
automatically and the corresponding position number appears on the remote con-
trol console.
The last set collimator positions, filter positions and dominants are stored auto-
matically assigned to each step position.
A maximum of 6 positions (5 steps) is possible.
After the programming run is completed, the X-ray system must be moved back
to the starting position (abdomen) (automatic stop).
Filling run
Initiate the individual steps of the X-ray system using the joystick or the key for
“X-ray system longitudinal movement” corresponding to the contrast medium
flow with monitoring on the live image monitor.
Selection
◆
Press this key.
– Peristepping is selected: the LED lights up.
Deselection
◆
Press this key or select another exposure system.
– Peristepping is deselected.
DR
DR
There is no automatic deselection of Peristepping after the end of the scene.
DR
DR
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...