Examination
Digital Radiography
AXIOM Iconos R200
AXD3-340.620.13.01.02
29
/ 52
Releasing the exposure
❏
To release the exposure, actuate the exposure release button in the unit re-
mote control console.
❏
The radiation ON indicators must light up during the exposure.
Radiation interruption
If you let go of the exposure release button during radiation, the exposure is in-
terrupted immediately and the tomographic movement is stopped.
Exposure end
Letting go of the exposure release button or elapsing of the tomographic time
ends the exposure.
Organ programs
All parameters necessary for the exposure are stored in organ programs.
➩
Refer to the FLUOROSPOT Compact and POLYDOROS SX operating instruc-
tions for further details on the organ programs and their programming.
Programs
The following programs can be stored:
– Tomography organ programs
Creation of programs
➩
Refer to the FLUOROSPOT Compact operating instructions.
Temporary changes of programs
For the following parameters stored in the organ program, it is possible to change
them temporarily with the corresponding operating elements:
– Film density
– Cu prefilter
Warning
If these radiation ON indicators light up although no switch has been actuated
and the current fluoroscopic image cannot be seen on the monitor:
◆
Switch the generator off immediately!
◆
Notify the SIEMENS Uptime Service.
◆
Discontinue operation of the system.
The changed parameters apply only until you select another / same organ pro-
gram or switch the system off.
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...