Technical Description
Technical Data
8
/ 12
AXD3-340.620.13.01.02
Operator Manual
Compression device
(optional)
Can be remotely controlled, cone radiotransparent, detachable and replaceable
(3 shapes)
Compression force:
5 to max. 155 N or 80 N
1
, movement blockage starting from 50 N compression
force
Speed:
Compression/decompression approx. 7.5 cm/s
Force indication:
Digital (LCD) in steps of 1 through 15 or 8
1
at the system remote control console
Cone/tabletop distance:
2 cm to 50 cm (standard cone)
Park position:
Cone in max. decompression position
Projection angle:
+ 30° to - 30°
Tomographic device
(optional)
Linear tomography in all table positions with cassette (only with cassette spotfilm
device) and digital (I.I.); automatic positioning of object and image receptor in the
working area.
Tomographic angle/
tomographic time
combinations:
8° with 0.4 s/ 0.8 s (working area 57.5 cm);
20° with 0.6 s/ 1.2 s (working area 42.5 cm);
40° with 1.2 s/ 2.5 s (working area 11.0 cm)
Tomographic angle at 300 mm fulcrum height
SID:
115 cm
Fulcrum height:
10 to 300 mm above tabletop, adjustable millimeter by millimeter at 0.5 cm/s;
Laser line light localizer (optional) for fulcrum height on the patient;
Operation on the system remote control console and table-side
Components
Image intensifier
SIRECON 33-4HDR
4 image fields:
∅
33 cm;
∅
22 cm;
∅
17 cm
; ∅
13 cm
SIRECON 40-4HDR
4 image fields:
∅
40 cm;
∅
30 cm;
∅
22 cm
; ∅
17 cm
TV system
VIDEOMED DHC
High-resolution TV camera with maintenance-free 1K CCD sensor for digital
fluoroscopy and fluororadiography
Dynamic response
max. 66 dB signal-to-noise ratio
TV matrix
Matrix 1024²
1
configuration for Japan only
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...