Examination
Digital Radiography
AXIOM Iconos R200
AXD3-340.620.13.01.02
31
/ 52
Releasing the exposure
❏
Release the exposure with the exposure release button on the system remote
control console or via the foot switch in the control room, or table-side (op-
tion).
❏
Radiation is released long as long as you continue to press the exposure re-
lease button.
❏
The radiation ON indicators must light up.
Radiation interruption
If you let go of the exposure release button during radiation, the exposure is in-
terrupted immediately.
Exposure end
Letting go of the exposure release button , a full image memory or reaching the
programmed maximum scene time ends the exposure series.
Examination procedure
Preparation
◆
Prepare the digital imaging system
◆
Move the tube assembly into the 0° position
◆
Position the patient
◆
Set a nominal source-image intensifier distance of 115 cm or 150 cm
◆
Bring the X-ray system and tabletop into the intended starting position.
You must ensure that the travel range for the X-ray system is sufficient.
◆
Press this key.
– The LED lights up: Periscanning is selected.
– Serialography is selected.
Warning
If these radiation ON indicators light up although no switch has been actuated
and the current fluoroscopic image cannot be seen on the monitor:
◆
Switch the system off immediately!
◆
Notify the SIEMENS Uptime Service.
◆
Discontinue operation of the system.
DR
DR
The LEDs in the Periscanning and start keys flash if the conditions for exposure
are not fulfilled.
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...