
System Overview
Protective Measures
AXIOM Iconos R200
AXD3-340.620.13.01.02
21
/ 44
◆
Do not stand at any place on the spotfilm device or on the covers of the table
support outside the marked areas provided for this.
– The covers can be deformed.
Components located underneath them are damaged and thus lead to oper-
ating disturbances.
◆
Place no objects on the operating areas of the control consoles and the table-
side control.
◆
In vertical table positions do not use the stand column, the tube assembly sup-
port arm, the tube assembly cover or the primary collimator as seat or support.
– This unallowed loading can lead to material breakage and damage to bear-
ings.
◆
Never put contrast medium cups or open containers with liquid or pasty con-
tents on the unit, on the remote console or on the control cabinets.
– Contrast medium can spill, leak or overflow into system parts and lead to
operational disturbances of the unit or to misinterpretation of exposures.
◆
When storing contrast medium in the cup holder on the compression carriage,
use only cups with a maximum volume of 0.25 liter made of unbreakable ma-
terials, i.e. under no circumstances glass or porcelain.
– Remove contrast medium traces immediately!
Measures for avoiding unwanted radiation
◆
Before starting system movements make sure that the foot switch for fluo-
roscopy and radiography (optional) in the examination room is not in the travel
range of the image intensifier light distribution system.
Caution
Unintentional activation of control elements for movements
Collision with patient, operator or equipment
◆
Do not load the remote console with any objects, accessories, folders or
documents.
To avoid unintentional activation of control elements for movements concerning
bed exposure, please adhere to the following workflow:
◆
Select aquisition mode
Bed exposure
at the generator control console.
◆
Tilt the system and move it to the correct SID.
◆
Rotate the tube assembly accordingly.
◆
Position the patient.
◆
Control patient and system.
◆
Don’t let patient stay in system area during absence of operator.
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...