POLYDOROS SX 65/80
Integrated Control Console
AXIOM Iconos R200
AXD3-340.620.13.01.02
15
/ 36
Fluoroscopic time limit
After a fluoroscopic time of 10 minutes
1
, fluoroscopy switches off automatically
and cannot be switched on again.
To continue working,
◆
let go of the fluoroscopy switch and actuate it again.
Exposure
Exposure release
You can release an exposure either with the release button on the generator ON/
OFF console
2
, on the system control console
3
or with the exposure release foot
switch in controlroom.
You can release an exposure if
❏
no error message appears, and
❏
none of the displays (kV, mAs, ms) flashes
The radiation indicator lights up and a signal sounds during the exposure release.
❏
Observe the following to increase the life of the tube assembly:
– If possible always select 80% tube load at the generator.
– It is essential that you program 80% tube load in the programs with small
focal spot and a tube voltage less than 70 kV.
– Work with the large focal spot below 70 kV tube voltage, if possible.
– Keep the prep time short and possibly “push through” to keep this time
short.
– Let go of the exposure release button immediately after the exposure.
1
At the wish of the customer, this parameter can be changed by SIEMENS Uptime Service, in accor-
dance with country-specific regulations.
2
Not possible in
DFR
and
Cassette
mode;
in
Vertix
mode with ACSS only when tube unit is rotated 180°
3
Not possible in
Bed
mode and
Vertix
mode
Check the selected exposure data on the integrated generator control console
before every radiation release.
The above instructions should be observed especially at workstations with high
patient throughput (accident stations) to counteract premature wear of the X-ray
tube assembly.
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...