Examination
Digital Radiography
32
/ 52
AXD3-340.620.13.01.02
Operator Manual
If the LED in the start key flashes,
◆
press the key.
– The compression device moves into park position, if present
– Oblique setting moves into the 0° position
Once these positions are reached, the flashing LED goes out.
◆
Select an organ program for Periscanning
◆
Position the patient and set the following parameters with the corresponding
operating elements:
– Primary collimation (rectangular diaphragm)
– Semitransparent wedge filters (if option is available)
– Cu prefilter (preferably filed in the organ program)
Adapting the
dominant
◆
Press one/several keys,
– The dominant preset in the organ program is overwritten.
Prepare the injector
◆
Connect the injector plug to the electronics cabinet (only with radiation-syn-
chronous injector release (option)).
◆
Set the injection data and switch to ready for injections.
Test run
The X-ray system stands in the starting position for the scanning process.
Start a test run without radiation.
◆
Switch the light localizer of the collimator on.
◆
Deflect the joystick up / down or the key to the left / right until the required
examination area has been scanned.
◆
If necessary correct the starting position.
Start
Position the patient so that the region the vascular filling of which should be im-
aged first is located in the central ray.
If a left / right dominant or both dominants are selected, the system switches
over automatically to the central circular dominant if the preselected dominant is
covered by the filters, since this would inevitably result in a faulty exposure.
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...