System Operation
System Settings
AXIOM Iconos R200
AXD3-340.620.13.01.02
13
/ 32
Behavior of the system settings in the case of a fault
If there is a fault while actuating one of the following movements
– tilting the table down/up
– moving the tabletop longitudinally
– moving the X-ray system transversely
then these movements are stopped and the fault is displayed as an error in the
message line of the integrated generator control console and stored.
Cancelling the fault:
◆
Briefly press the exposure release button on the generator ON/OFF console
up to the 1st pressure point and let go again. Don’t press more!.
– The fault is cancelled as a rule
Actuating the same movement again
If a fault occurs again when the same movement is actuated, then these move-
ments are once again stopped and the fault is displayed as an error in the mes-
sage line of the integrated generator control console and stored.
Cancelling the fault:
◆
Briefly press the exposure release button on the generator ON/OFF console
up to the 1st pressure point and let go again. Don’t press more!.
– The fault is cancelled as a rule by this
Actuating the same movement again
If the same movement is actuated again and there is a fault once again, then a
warning signal (whistle) sounds and this movement is blocked. The following er-
ror message is displayed in the message line of the integrated generator control
console:
“Severe error, please call service”
All other movements are possible only step by step. A warning signal (whistle)
sounds on actuating and during the movement.
◆
Rescue the patient.
◆
Call the Siemens Uptime Service
Warning
Collision monitoring is inoperable.
The operator is responsible for avoiding collisions of the system with objects
and the floor.
◆
Actuate system movements only with direct visual control.
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...