System Overview
Protective Measures
18
/ 44
AXD3-340.620.13.01.02
Operator Manual
These objects are not detected by the collision monitoring of the unit. Avoiding
collisions of the unit with these objects is subject to the operator’s duty to take
care.
If a collision of the unit with a rigid obstacle (e.g. step) has occurred,
◆
press one of the red emergency stop buttons,
◆
rescue the patient,
◆
immediately notify the SIEMENS Uptime Service.
❏
Avoid standing or sitting immediately adjacent to the system and especially
do not sit next to the system with your legs or knees under the cross-beam at
the head or foot end of the table.
❏
Take care that during system movements no one is in the area between the
unit base and table.
❏
Take care that with the footboard attached there is a risk of collision with the
extended cone (optional) when the tabletop and / or the longitudinal carriage
are moving.
❏
Do not grasp in the loading shaft of the spotfilm device because of the risk of
crushing.
Patient positioning
❏
All safety-related equipment must be installed and operable. In particular the
handgrip strips (head end and lateral), handgrip, footboard, foot restraints,
compression belt and shoulder supports.
❏
The patient’s hands, arms, legs, head and hair must not extend unsecured be-
yond the edge of the tabletop.
❏
Observe the patient while moving the tabletop and in system movements and
take care that any catheter is correctly located.
❏
In examinations with the table tilted up vertically, the footboard serves as an
adjustable step or seat.
– Ensure that the footboard is locked together with the tabletop on both
sides.
– Check the firm location of the footboard.
Warning
Under no circumstances must the unit be tilted down further or tilted up, since
externally invisible, but safety-relevant damage to the tilting drive may occur.
Severe consequential damage including personal damage cannot be excluded
in this case.
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...