Examination
Digital Radiography
AXIOM Iconos R200
AXD3-340.620.13.01.02
33
/ 52
Scanning process
The X-ray system stands in the starting position for the scanning process.
◆
Press an exposure release button and keep it pressed up to the end of the ex-
amination, i.e. until the scanning process is ended.
◆
Trigger the injection, automatically or manually
◆
Deflect the joystick up / down or press the key to the left / right until the entire
examination area has been scanned.
◆
If necessary reduce the frame rate in a maximum of 2 steps, as filed in the
organ program (saving dose and storage capacity).
End of the scene
◆
Let the exposure release button go.
Organ programs
All parameters necessary for the exposure are stored in organ programs.
➩
Refer to the FLUOROSPOT Compact Operator Manual and the POLYDOROS
SX register for further details on the organ programs and their programming.
Creation of programs
➩
Refer to the FLUOROSPOT Compact Operator Manual.
Temporary changes of programs
For the following parameters stored in the organ program, it is possible to change
them temporarily with the corresponding operating elements:
– Film density
– Dominants
– Cu prefilter
– Automatic level 1, 2, 3
Follow the contrast medium flow on the monitor.
Adapt the movement speed of the X-ray system to the contrast medium flow.
The exposure series is also ended if the image memory is full and the maximum
scene time has been reached, independently of the position in which the system
is located.
The changed parameters apply only until you select another / same organ pro-
gram or switch the system off.
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...